DeltaVision softWoRx 7.2.1 Update May 2020 This update is for systems currently running 7.X (or 7.2.1 pre-release) updating to the latest version of 7.2.1. Prerequisites * You must have at least softWoRx 7.X installed * You must have CentOS 5 or higher installed * You must have access to the Linux workstation's "root" password md5sum: 5d86ba1e8380dd9469ae74b1704e3e20 dv721_update.tar This file should be downloaded or transferred to the desktop of a user account on the workstation. Install: 1. Shut down any DeltaVision windows on the acquisition workstation. 2. Right click the icon named dv721_update.tar (on the workstation's Desktop background) and choose Extract Here from the menu. 3. A folder called dv721_update will be created. 4. If the extraction does not result in a window being displayed, double-click on the dv721_update folder icon to display its contents. 5. In that folder, click the icon Click_To_Update.sh and follow the instructions. This will ask for root password, make backup copies of the current softWoRx folders, then apply the updates, including sending an update to the IC/MIC (if attached.) The remaining steps update the software on the IC/MIC--stand-alone workstations or Ultra systems do not need to continue. For all IC/MIC systems: 10. To finalize the firmware update on the IC/MIC, press the power button to shut it down. 11. When the IC/MIC has powered off, press the power button again and wait for the IC/MIC to power up, update itself, and then automatically power off again. 12. Once the IC/MIC has automatically shut itself off after updating its firmware, push the IC/MIC power button again to power it on for use.